The same device (matched on an identical UDI-DI) is registered in other jurisdictions. EU coverage (EUDAMED) is still phasing in — absence here does not mean a device is not marketed there.

  • European UnionVAPROX0724995MMVAPROXHCSTRL6L
Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

9 linked reportsMalfunction 9
13452019-01: 0 reports for this device; peer average 3/device2019-02: 0 reports for this device; peer average 2.83/device2019-03: 0 reports for this device; peer average 3/device2019-04: 0 reports for this device; peer average 2/device2019-05: 0 reports for this device; peer average 1.4/device2019-06: 0 reports for this device; peer average 1.4/device2019-07: 0 reports for this device; peer average 3.67/device2019-08: 0 reports for this device; peer average 3/device2019-09: 0 reports for this device; peer average 2.4/device2019-10: 0 reports for this device; peer average 3.8/device2019-11: 0 reports for this device; peer average 3/device2019-12: 0 reports for this device; peer average 4.33/device2020-01: 0 reports for this device; peer average 2.17/device2020-02: 0 reports for this device; peer average 4/device2020-03: 0 reports for this device; peer average 1.71/device2020-04: 0 reports for this device; peer average 3.14/device2020-05: 0 reports for this device; peer average 1.8/device2020-06: 0 reports for this device; peer average 3/device2020-07: 0 reports for this device; peer average 2.29/device2020-08: 0 reports for this device; peer average 2/device2020-09: 0 reports for this device; peer average 2.5/device2020-10: 0 reports for this device; peer average 2.33/device2020-11: 0 reports for this device; peer average 2.5/device2020-12: 0 reports for this device; peer average 1.6/device2021-01: 0 reports for this device; peer average 2.33/device2021-02: 0 reports for this device; peer average 3.5/device2021-03: 0 reports for this device; peer average 2.43/device2021-04: 0 reports for this device; peer average 4.5/device2021-05: 0 reports for this device; peer average 2/device2021-06: 0 reports for this device; peer average 2/device2021-07: 0 reports for this device; peer average 2.83/device2021-08: 0 reports for this device; peer average 2.08/device2021-09: 0 reports for this device; peer average 2.8/device2021-10: 0 reports for this device; peer average 2.13/device2021-11: 0 reports for this device; peer average 2.13/device2021-12: 0 reports for this device; peer average 2.57/device2022-01: 0 reports for this device; peer average 2.5/device2022-02: 0 reports for this device; peer average 1.17/device2022-03: 0 reports for this device; peer average 3/device2022-04: 0 reports for this device; peer average 2.4/device2022-05: 0 reports for this device; peer average 2.88/device2022-06: 0 reports for this device; peer average 1.83/device2022-07: 0 reports for this device; peer average 2.67/device2022-08: 0 reports for this device; peer average 2/device2022-09: 0 reports for this device; peer average 1.86/device2022-10: 0 reports for this device; peer average 1.63/device2022-11: 0 reports for this device; peer average 1.88/device2022-12: 0 reports for this device; peer average 3.17/device2023-01: 0 reports for this device; peer average 2/device2023-02: 0 reports for this device; peer average 1.73/device2023-03: 0 reports for this device; peer average 1.78/device2023-04: 0 reports for this device; peer average 2.5/device2023-05: 0 reports for this device; peer average 2.8/device2023-06: 0 reports for this device; peer average 1.92/device2023-07: 0 reports for this device; peer average 2.3/device2023-08: 0 reports for this device; peer average 2.4/device2023-09: 0 reports for this device; peer average 1.56/device2023-11: 0 reports for this device; peer average 1.5/device2023-12: 0 reports for this device; peer average 2.75/device2024-01: 0 reports for this device; peer average 1.33/device2024-03: 1 reports for this device; peer average 1.25/device2024-04: 0 reports for this device; peer average 1/device2024-05: 0 reports for this device; peer average 1.5/device2024-06: 0 reports for this device; peer average 2/device2024-08: 0 reports for this device; peer average 3.67/device2024-09: 0 reports for this device; peer average 4.75/device2024-10: 0 reports for this device; peer average 2.5/device2024-11: 0 reports for this device; peer average 1.7/device2024-12: 2 reports for this device; peer average 1.25/device2025-01: 2 reports for this device; peer average 1.6/device2025-02: 0 reports for this device; peer average 1.67/device2025-03: 0 reports for this device; peer average 2/device2025-04: 0 reports for this device; peer average 2.75/device2025-05: 0 reports for this device; peer average 1.75/device2025-06: 2 reports for this device; peer average 4.6/device2025-07: 0 reports for this device; peer average 5/device2025-08: 0 reports for this device; peer average 3.33/device2025-09: 0 reports for this device; peer average 2.67/device2025-10: 0 reports for this device; peer average 2.33/device2025-11: 0 reports for this device; peer average 3/device2025-12: 2 reports for this device; peer average 1.71/device2026-01: 0 reports for this device; peer average 3.4/device2026-02: 0 reports for this device; peer average 5/device2026-03: 0 reports for this device; peer average 2.5/device2026-04: 0 reports for this device; peer average 2/device2026-05: 0 reports for this device; peer average 2/device2026-06: 0 reports for this device; peer average 2.8/device010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809111201030405060809101112010203040506070809101112010203040506
Monthly adverse-event reports for this device versus the product-code average.
MonthThis device (reports)Product-code average (reports per device)
2019-0103
2019-0202.83
2019-0303
2019-0402
2019-0501.4
2019-0601.4
2019-0703.67
2019-0803
2019-0902.4
2019-1003.8
2019-1103
2019-1204.33
2020-0102.17
2020-0204
2020-0301.71
2020-0403.14
2020-0501.8
2020-0603
2020-0702.29
2020-0802
2020-0902.5
2020-1002.33
2020-1102.5
2020-1201.6
2021-0102.33
2021-0203.5
2021-0302.43
2021-0404.5
2021-0502
2021-0602
2021-0702.83
2021-0802.08
2021-0902.8
2021-1002.13
2021-1102.13
2021-1202.57
2022-0102.5
2022-0201.17
2022-0303
2022-0402.4
2022-0502.88
2022-0601.83
2022-0702.67
2022-0802
2022-0901.86
2022-1001.63
2022-1101.88
2022-1203.17
2023-0102
2023-0201.73
2023-0301.78
2023-0402.5
2023-0502.8
2023-0601.92
2023-0702.3
2023-0802.4
2023-0901.56
2023-1101.5
2023-1202.75
2024-0101.33
2024-0311.25
2024-0401
2024-0501.5
2024-0602
2024-0803.67
2024-0904.75
2024-1002.5
2024-1101.7
2024-1221.25
2025-0121.6
2025-0201.67
2025-0302
2025-0402.75
2025-0501.75
2025-0624.6
2025-0705
2025-0803.33
2025-0902.67
2025-1002.33
2025-1103
2025-1221.71
2026-0103.4
2026-0205
2026-0302.5
2026-0402
2026-0502
2026-0602.8
This device (reports/month)Average device in MLR (reports/device/month)
Failure-mode summaryAI-generated

9 linked report narratives available — no AI summary generated yet.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2018-07-26Z-2959-2018Class IITerminatedThere is the potential for a scenario where a punctured cup of sterilant expires while the system is powered down, upon power-up the unit will not detect the expired cup of sterilant; allowing for subsequent cycles to be run.Distribution: Worldwide Distribution: US(nationwide) and to countries of: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Belarus, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Malaysia, Malta, Mexico, Morocco, Nepal, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Suriname, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.2889 unitsclearance family
2017-09-28Z-0772-2018Class IITerminatedCorrection to update sterilizer software to prohibit the use of expired sterilant prior to the start of a sterilization cycle.Distribution: Distributed in 44 states: AL, AK, AZ, AR, CA, CO, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY; Puerto Rico, and the following countries: Algeria, Australia, Azerbaijan, Bahrain, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Republic of Korea, Kuwait, Latvia, Malaysia, Mexico, Morocco, Myanmar, Panama, Peru, Philippines, Poland, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, and Vietnam.406clearance family
2011-08-01Z-3041-2011TerminatedSTERIS learned that Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems users may experience premature wear of the vacuum pump and filtering assembly components. Users may also experience incomplete aeration of the device's chamber following a failed leak test.clearance family
Individual MAUDE reports
DateReport #TypeNarrativeLink basis
2025-12-173003950207-2025-00012MalfunctionAvailablecatalog number
2025-06-123003950207-2025-00007MalfunctionAvailablecatalog number
2025-01-023003950207-2025-00001MalfunctionAvailablecatalog number
2024-12-313003950207-2024-00011MalfunctionAvailablecatalog number
2024-03-283003950207-2024-00003MalfunctionAvailableexact di

Showing the most recent 5 of 9 linked reports.

Regulatory clearances
  • 510(k)K233065Substantially equivalent (SESE) · 2023-10-20clearance family
  • 510(k)K223476Substantially equivalent (SESE) · 2023-08-07clearance family
  • 510(k)K222093Substantially equivalent (SESE) · 2022-09-09clearance family
  • 510(k)K190917Substantially equivalent (SESE) · 2019-05-09clearance family
  • 510(k)K190103Substantially equivalent (SESE) · 2019-04-05clearance family
  • 510(k)K182568Substantially equivalent (SESE) · 2019-01-03clearance family
  • 510(k)K172754Substantially equivalent (SESE) · 2018-02-09clearance family
  • 510(k)K172319Substantially equivalent (SESE) · 2018-01-18clearance family
  • 510(k)K162413Substantially equivalent (SESE) · 2017-03-31clearance family
  • 510(k)K160433Substantially equivalent (SESE) · 2016-07-06clearance family
  • 510(k)K140498Substantially equivalent (SESE) · 2014-07-24clearance family
  • 510(k)K131120Substantially equivalent (SESE) · 2013-07-16clearance family
  • 510(k)K120632Substantially equivalent (SESE) · 2012-06-26clearance family
  • 510(k)K112813Substantially equivalent (SESE) · 2011-12-01clearance family
  • 510(k)K112760Substantially equivalent (SESE) · 2011-11-22clearance family
  • 510(k)K111810Substantially equivalent (SESE) · 2011-09-02clearance family
  • 510(k)K102394Substantially equivalent (SESE) · 2011-08-18clearance family
  • 510(k)K102330Substantially equivalent (SESE) · 2011-08-12clearance family

Predicate lineage

Cited as predicate by: K120632, K131120, K140498, K160433, K162413, K172319

Cited as predicate by: K160433, K162413, K172319

K190103

Cited as predicate by: K190027, K190917, K222093, K222543

Cited as predicate by: K223476

Cited as predicate by: K233065

Lineage each submission cites, walked transitively (indented = the cited device's own predicates). Extracted from public 510(k) summary text — heuristic, and includes reference devices mentioned alongside true predicates.

Manufacturer compliance record
DateTypeOutcomeFacilityCitations
2026-04-16NAINo action indicatedSteris Corporation0
2026-04-16NAINo action indicatedSteris Corporation0
2026-04-03NAINo action indicatedIsomedix Operations Inc.0
2026-03-12NAINo action indicatedIsomedix Operations Inc0
2026-02-26VAIVoluntary action indicatedSteris S.p.A.1
2026-02-12VAIVoluntary action indicatedSynergy Health Ede BV2
2026-02-12NAINo action indicatedSynergy Health Marseille0
2025-12-12OAIOfficial action indicatedBlack Diamond Video Inc.3
2025-08-01VAIVoluntary action indicatedSTERIS Laboratories, Inc.2
2025-06-18NAINo action indicatedIsomedix Operations Inc.0
2025-06-17NAINo action indicatedIsomedix Operations Inc0
2025-06-05NAINo action indicatedIsomedix Operations, Inc.0
2025-04-01NAINo action indicatedSteris Corporation0
2025-02-21NAINo action indicatedIsomedix Operations Inc.0
2025-01-23NAINo action indicatedSynergy Health Allershausen GmbH0
2024-12-12VAIVoluntary action indicatedIsomedix Operations, Inc0
2024-12-05NAINo action indicatedSteris (Synergy Health Ede BV)0
2024-11-21VAIVoluntary action indicatedSteris Applied Sterilization Technologies ULC2
2024-09-23NAINo action indicatedIsomedix Operations, Inc.0
2024-09-19NAINo action indicatedIsomedix Operations, Inc.0
2015-02-20WLWarning LetterKey Surgical LLC
2014-05-22WLWarning LetterIsomedix Operations, Inc.
2012-05-29WLWarning LetterAlbert Browne Ltd.
2011-02-09WLWarning LetterSteris Corporation
2011-02-09WLWarning LetterSteris Corporation
2010-09-27WLWarning LetterIsomedix Operations, Inc.
2010-04-20INJInjunctionSteris Corporation
2010-04-20INJInjunctionSteris Corporation
2009-07-31WLWarning LetterMedivators Inc