Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

No adverse-event reports are linked to this device.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2018-07-26Z-2959-2018Class IITerminatedThere is the potential for a scenario where a punctured cup of sterilant expires while the system is powered down, upon power-up the unit will not detect the expired cup of sterilant; allowing for subsequent cycles to be run.Distribution: Worldwide Distribution: US(nationwide) and to countries of: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Belarus, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Malaysia, Malta, Mexico, Morocco, Nepal, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Suriname, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.2889 unitscode info di
2017-09-28Z-0772-2018Class IITerminatedCorrection to update sterilizer software to prohibit the use of expired sterilant prior to the start of a sterilization cycle.Distribution: Distributed in 44 states: AL, AK, AZ, AR, CA, CO, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY; Puerto Rico, and the following countries: Algeria, Australia, Azerbaijan, Bahrain, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Republic of Korea, Kuwait, Latvia, Malaysia, Mexico, Morocco, Myanmar, Panama, Peru, Philippines, Poland, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, and Vietnam.406clearance family
2011-08-01Z-3041-2011TerminatedSTERIS learned that Amsco V-PRO 1 and V-PRO 1 Plus Low Temperature Sterilization Systems users may experience premature wear of the vacuum pump and filtering assembly components. Users may also experience incomplete aeration of the device's chamber following a failed leak test.clearance family
Regulatory clearances
  • 510(k)K190103Substantially equivalent (SESE) · 2019-04-05clearance family
  • 510(k)K172319Substantially equivalent (SESE) · 2018-01-18clearance family
  • 510(k)K162413Substantially equivalent (SESE) · 2017-03-31clearance family
  • 510(k)K160433Substantially equivalent (SESE) · 2016-07-06clearance family
  • 510(k)K131120Substantially equivalent (SESE) · 2013-07-16clearance family
  • 510(k)K120632Substantially equivalent (SESE) · 2012-06-26clearance family
  • 510(k)K112813Substantially equivalent (SESE) · 2011-12-01clearance family
  • 510(k)K111810Substantially equivalent (SESE) · 2011-09-02clearance family
  • 510(k)K102394Substantially equivalent (SESE) · 2011-08-18clearance family
  • 510(k)K102330Substantially equivalent (SESE) · 2011-08-12clearance family

Predicate lineage

Cited as predicate by: K120632, K131120, K140498, K160433, K162413, K172319

Cited as predicate by: K160433, K162413, K172319

K190103

Cited as predicate by: K190027, K190917, K222093, K222543

Lineage each submission cites, walked transitively (indented = the cited device's own predicates). Extracted from public 510(k) summary text — heuristic, and includes reference devices mentioned alongside true predicates.

Manufacturer compliance record

None in the loaded window.

Similar devices
Select up to two to place side-by-side with this device.