Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

No adverse-event reports are linked to this device.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2025-07-07Z-2273-2025Class IOpen, ClassifiedCertain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the guidewire or embolic protection device (EPD) after successful stent deployment. The most serious potential adverse health consequence is stroke. Other risks include delay to procedure, vessel injury, vessel spasm, or stent disruption/damage necessitating additional intervention.Distribution: Domestic US Nationwide distribution including Puerto Rico. International distribution worldwide.26,570 (1,333 US; 25,174 OUS)code info di
Regulatory clearances
  • PMAP050019Approved (APPR) · 2025-09-19 · +19 earlier decisionsclearance family
Manufacturer compliance record
DateTypeOutcomeFacilityCitations
2025-04-07OAIOfficial action indicatedBoston Scientific Corporation9
2025-04-07OAIOfficial action indicatedBoston Scientific Corporation9
2025-03-18NAINo action indicatedBoston Scientific Corporation0
2025-03-18NAINo action indicatedBoston Scientific Corporation0
2024-12-06NAINo action indicatedNalu Medical, Inc.0
2024-11-26VAIVoluntary action indicatedBoston Scientific Corporation3
2024-11-26VAIVoluntary action indicatedBoston Scientific Corporation3
2024-06-13NAINo action indicatedElutia Inc0
2024-01-23VAIVoluntary action indicatedSilk Road Medical1
2023-12-05NAINo action indicatedValencia Technologies Corporation0
2023-12-01VAIVoluntary action indicatedCorbin Clinical Resources, Llc1
2023-08-01NAINo action indicatedBoston Scientific Corporation0
2023-08-01NAINo action indicatedBoston Scientific Corporation0
2023-04-05NAINo action indicatedIntera Oncology, Inc.0
2022-07-28NAINo action indicatedBoston Scientific Corporation0
2022-07-28NAINo action indicatedBoston Scientific Corporation0
2022-06-02VAIVoluntary action indicatedBoston Scientific de Costa Rica SRL0
2022-03-23NAINo action indicatedAxonics, Inc.0
2021-11-17VAIVoluntary action indicatedValencia Technologies Corporation1
2021-10-29NAINo action indicatedBoston Scientific Corporation0
2014-04-10WLWarning LetterBoston Scientific Corporation
Similar devices