Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

287 linked reportsMalfunction 283Injury 4
4711142019-01: 0 reports for this device; peer average 2.42/device2019-02: 0 reports for this device; peer average 3.53/device2019-03: 0 reports for this device; peer average 2.95/device2019-04: 1 reports for this device; peer average 3.13/device2019-05: 0 reports for this device; peer average 2.46/device2019-06: 1 reports for this device; peer average 3.58/device2019-07: 0 reports for this device; peer average 2.42/device2019-08: 0 reports for this device; peer average 2.47/device2019-09: 11 reports for this device; peer average 2.47/device2019-10: 1 reports for this device; peer average 2.71/device2019-11: 0 reports for this device; peer average 2.12/device2019-12: 0 reports for this device; peer average 2.35/device2020-01: 5 reports for this device; peer average 2.55/device2020-02: 0 reports for this device; peer average 2.7/device2020-03: 0 reports for this device; peer average 2.28/device2020-04: 0 reports for this device; peer average 2.54/device2020-05: 0 reports for this device; peer average 1.98/device2020-06: 1 reports for this device; peer average 2.78/device2020-07: 1 reports for this device; peer average 2.77/device2020-08: 1 reports for this device; peer average 4.33/device2020-09: 0 reports for this device; peer average 4.4/device2020-10: 0 reports for this device; peer average 3.99/device2020-11: 0 reports for this device; peer average 4.47/device2020-12: 0 reports for this device; peer average 3.87/device2021-01: 0 reports for this device; peer average 3.99/device2021-02: 3 reports for this device; peer average 3.89/device2021-03: 0 reports for this device; peer average 4.83/device2021-04: 2 reports for this device; peer average 4.7/device2021-05: 2 reports for this device; peer average 4.05/device2021-06: 3 reports for this device; peer average 4.28/device2021-07: 2 reports for this device; peer average 3.88/device2021-08: 0 reports for this device; peer average 3.87/device2021-09: 2 reports for this device; peer average 3.91/device2021-10: 4 reports for this device; peer average 3.99/device2021-11: 5 reports for this device; peer average 3.49/device2021-12: 10 reports for this device; peer average 4.46/device2022-01: 4 reports for this device; peer average 4.57/device2022-02: 0 reports for this device; peer average 5.03/device2022-03: 4 reports for this device; peer average 5.45/device2022-04: 4 reports for this device; peer average 6.48/device2022-05: 2 reports for this device; peer average 5.46/device2022-06: 4 reports for this device; peer average 6.07/device2022-07: 3 reports for this device; peer average 6.34/device2022-08: 0 reports for this device; peer average 6.38/device2022-09: 4 reports for this device; peer average 5.53/device2022-10: 4 reports for this device; peer average 5.35/device2022-11: 10 reports for this device; peer average 5.52/device2022-12: 4 reports for this device; peer average 6/device2023-01: 12 reports for this device; peer average 5.73/device2023-02: 5 reports for this device; peer average 4.87/device2023-03: 2 reports for this device; peer average 5.86/device2023-04: 7 reports for this device; peer average 5.47/device2023-05: 10 reports for this device; peer average 5.86/device2023-06: 10 reports for this device; peer average 7.1/device2023-07: 4 reports for this device; peer average 6.1/device2023-08: 6 reports for this device; peer average 6.28/device2023-09: 6 reports for this device; peer average 5.92/device2023-10: 4 reports for this device; peer average 6.38/device2023-11: 2 reports for this device; peer average 5.76/device2023-12: 10 reports for this device; peer average 6.05/device2024-01: 2 reports for this device; peer average 5.53/device2024-02: 14 reports for this device; peer average 6.18/device2024-03: 6 reports for this device; peer average 5.56/device2024-04: 4 reports for this device; peer average 6.32/device2024-05: 7 reports for this device; peer average 5.3/device2024-06: 6 reports for this device; peer average 7.07/device2024-07: 1 reports for this device; peer average 5.75/device2024-08: 1 reports for this device; peer average 6.5/device2024-09: 1 reports for this device; peer average 6.28/device2024-10: 4 reports for this device; peer average 7.11/device2024-11: 2 reports for this device; peer average 4.55/device2024-12: 5 reports for this device; peer average 4.77/device2025-01: 1 reports for this device; peer average 4.42/device2025-02: 3 reports for this device; peer average 4.03/device2025-03: 3 reports for this device; peer average 5.17/device2025-04: 4 reports for this device; peer average 4.67/device2025-05: 1 reports for this device; peer average 4.69/device2025-06: 6 reports for this device; peer average 4.58/device2025-07: 2 reports for this device; peer average 4.68/device2025-08: 6 reports for this device; peer average 5.45/device2025-09: 0 reports for this device; peer average 3.97/device2025-10: 4 reports for this device; peer average 4.59/device2025-11: 1 reports for this device; peer average 4.86/device2025-12: 3 reports for this device; peer average 6.15/device2026-01: 4 reports for this device; peer average 4.95/device2026-02: 2 reports for this device; peer average 5.05/device2026-03: 5 reports for this device; peer average 5.77/device2026-04: 7 reports for this device; peer average 6.2/device2026-05: 1 reports for this device; peer average 5.86/device2026-06: 5 reports for this device; peer average 5.01/device010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506
Monthly adverse-event reports for this device versus the product-code average.
MonthThis device (reports)Product-code average (reports per device)
2019-0102.42
2019-0203.53
2019-0302.95
2019-0413.13
2019-0502.46
2019-0613.58
2019-0702.42
2019-0802.47
2019-09112.47
2019-1012.71
2019-1102.12
2019-1202.35
2020-0152.55
2020-0202.7
2020-0302.28
2020-0402.54
2020-0501.98
2020-0612.78
2020-0712.77
2020-0814.33
2020-0904.4
2020-1003.99
2020-1104.47
2020-1203.87
2021-0103.99
2021-0233.89
2021-0304.83
2021-0424.7
2021-0524.05
2021-0634.28
2021-0723.88
2021-0803.87
2021-0923.91
2021-1043.99
2021-1153.49
2021-12104.46
2022-0144.57
2022-0205.03
2022-0345.45
2022-0446.48
2022-0525.46
2022-0646.07
2022-0736.34
2022-0806.38
2022-0945.53
2022-1045.35
2022-11105.52
2022-1246
2023-01125.73
2023-0254.87
2023-0325.86
2023-0475.47
2023-05105.86
2023-06107.1
2023-0746.1
2023-0866.28
2023-0965.92
2023-1046.38
2023-1125.76
2023-12106.05
2024-0125.53
2024-02146.18
2024-0365.56
2024-0446.32
2024-0575.3
2024-0667.07
2024-0715.75
2024-0816.5
2024-0916.28
2024-1047.11
2024-1124.55
2024-1254.77
2025-0114.42
2025-0234.03
2025-0335.17
2025-0444.67
2025-0514.69
2025-0664.58
2025-0724.68
2025-0865.45
2025-0903.97
2025-1044.59
2025-1114.86
2025-1236.15
2026-0144.95
2026-0225.05
2026-0355.77
2026-0476.2
2026-0515.86
2026-0655.01
This device (reports/month)Average device in GCJ (reports/device/month)
Failure-mode summaryAI-generated

40 linked report narratives available — no AI summary generated yet.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2022-01-14Z-0728-2022Class IIOpen, ClassifiedA software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.Distribution: Worldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL, IN, FR, ES, HK, IT, GB, CO, PE, MY, TH, CA, JP, MX, AR, NP, SG, PH, ZA, CL, KR, CN11,955code info di
Individual MAUDE reports
DateReport #TypeNarrativeLink basis
2026-06-290002936485-2026-00417Malfunctionexact di
2026-06-290002936485-2026-00418MalfunctionAvailableexact di
2026-06-260002936485-2026-00415MalfunctionAvailableexact di
2026-06-030002936485-2026-00358MalfunctionAvailableexact di
2026-06-010002936485-2026-00351MalfunctionAvailableexact di
2026-05-200002936485-2026-00307MalfunctionAvailableexact di
2026-04-220002936485-2026-00255MalfunctionAvailableexact di
2026-04-080002936485-2026-00227MalfunctionAvailableexact di
2026-04-030002936485-2026-00222MalfunctionAvailableexact di
2026-04-030002936485-2026-00221MalfunctionAvailableexact di
2026-04-010002936485-2026-00203MalfunctionAvailableexact di
2026-04-010002936485-2026-00204MalfunctionAvailableexact di
2026-04-010002936485-2026-00200MalfunctionAvailableexact di
2026-03-260002936485-2026-00187MalfunctionAvailableexact di
2026-03-100002936485-2026-00153MalfunctionAvailableexact di

Showing the most recent 200 of 287 linked reports.

Showing 115 of 200
Regulatory clearances

None linked.

Manufacturer compliance record
DateTypeOutcomeFacilityCitations
2026-06-18NAINo action indicatedInternational Life Sciences0
2026-02-20VAIVoluntary action indicatedStryker Puerto Rico, LLC2
2025-12-11NAINo action indicatedStryker Manufacturing Kalamazoo0
2025-05-22NAINo action indicatedStryker Macroom0
2025-03-27VAIVoluntary action indicatedStryker GmbH1
2025-03-27VAIVoluntary action indicatedStryker GmbH1
2025-02-27NAINo action indicatedStryker Leibinger GmbH & Co. KG0
2025-01-30NAINo action indicatedSerf0
2024-12-18VAIVoluntary action indicatedStryker Neurovascular1
2024-11-15VAIVoluntary action indicatedStryker Medical Division of Stryker Corporation2
2024-11-13NAINo action indicatedPOM Medical LLC0
2024-07-25VAIVoluntary action indicatedStryker Ireland, Ltd., Instruments Division1
2024-05-03NAINo action indicatedStryker Neurovascular, Inc.0
2024-04-18NAINo action indicatedPhysio-Control, Inc.0
2023-12-01VAIVoluntary action indicatedHowmedica Osteonics Corp.1
2023-05-15VAIVoluntary action indicatedLimFlow, Inc.2
2023-04-25NAINo action indicatedStryker Puerto Rico, LLC0
2023-01-24VAIVoluntary action indicatedStryker Corporation2
2023-01-24VAIVoluntary action indicatedStryker Corporation2
2022-03-08VAIVoluntary action indicatedPOM Medical LLC0
2017-04-06WLWarning LetterEntellus Medical Inc.
2015-04-08WLWarning LetterThermedx LLC
2013-11-26INJInjunctionPhysio-Control, Inc.
2013-03-06WLWarning LetterStryker Corporation
2013-03-06WLWarning LetterStryker Corporation