Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

97 linked reportsMalfunction 97
214263842019-01: 0 reports for this device; peer average 6.86/device2019-02: 0 reports for this device; peer average 4.35/device2019-03: 0 reports for this device; peer average 4.42/device2019-04: 0 reports for this device; peer average 6.12/device2019-05: 0 reports for this device; peer average 4.8/device2019-06: 0 reports for this device; peer average 5.04/device2019-07: 0 reports for this device; peer average 5.86/device2019-08: 0 reports for this device; peer average 6.1/device2019-09: 0 reports for this device; peer average 4.99/device2019-10: 0 reports for this device; peer average 5.85/device2019-11: 0 reports for this device; peer average 5.62/device2019-12: 0 reports for this device; peer average 5.41/device2020-01: 0 reports for this device; peer average 4.81/device2020-02: 0 reports for this device; peer average 5.57/device2020-03: 0 reports for this device; peer average 5.68/device2020-04: 0 reports for this device; peer average 3.57/device2020-05: 0 reports for this device; peer average 3.76/device2020-06: 0 reports for this device; peer average 5.14/device2020-07: 0 reports for this device; peer average 6.56/device2020-08: 0 reports for this device; peer average 4.64/device2020-09: 0 reports for this device; peer average 5.86/device2020-10: 0 reports for this device; peer average 5.55/device2020-11: 0 reports for this device; peer average 5.14/device2020-12: 0 reports for this device; peer average 5.31/device2021-01: 0 reports for this device; peer average 5.17/device2021-02: 0 reports for this device; peer average 4.89/device2021-03: 0 reports for this device; peer average 5.7/device2021-04: 0 reports for this device; peer average 5.26/device2021-05: 0 reports for this device; peer average 5.14/device2021-06: 0 reports for this device; peer average 5.83/device2021-07: 0 reports for this device; peer average 4.89/device2021-08: 0 reports for this device; peer average 4.3/device2021-09: 0 reports for this device; peer average 4.77/device2021-10: 0 reports for this device; peer average 6.17/device2021-11: 0 reports for this device; peer average 5.57/device2021-12: 0 reports for this device; peer average 5.56/device2022-01: 0 reports for this device; peer average 5.04/device2022-02: 0 reports for this device; peer average 5.15/device2022-03: 0 reports for this device; peer average 5.75/device2022-04: 0 reports for this device; peer average 4.98/device2022-05: 0 reports for this device; peer average 5.78/device2022-06: 0 reports for this device; peer average 6.3/device2022-07: 0 reports for this device; peer average 4.36/device2022-08: 1 reports for this device; peer average 4.59/device2022-09: 0 reports for this device; peer average 5/device2022-10: 1 reports for this device; peer average 4.76/device2022-11: 0 reports for this device; peer average 4.61/device2022-12: 0 reports for this device; peer average 4.04/device2023-01: 0 reports for this device; peer average 5.52/device2023-02: 2 reports for this device; peer average 4.59/device2023-03: 0 reports for this device; peer average 4.87/device2023-04: 2 reports for this device; peer average 4.78/device2023-05: 2 reports for this device; peer average 5.14/device2023-06: 1 reports for this device; peer average 4.78/device2023-07: 0 reports for this device; peer average 4.82/device2023-08: 1 reports for this device; peer average 4.44/device2023-09: 2 reports for this device; peer average 5.23/device2023-10: 84 reports for this device; peer average 5.1/device2023-11: 0 reports for this device; peer average 4.38/device2023-12: 1 reports for this device; peer average 4.79/device2024-01: 0 reports for this device; peer average 5.54/device2024-02: 0 reports for this device; peer average 5.46/device2024-03: 0 reports for this device; peer average 6.16/device2024-04: 0 reports for this device; peer average 6.04/device2024-05: 0 reports for this device; peer average 6.28/device2024-06: 0 reports for this device; peer average 5.39/device2024-07: 0 reports for this device; peer average 5.05/device2024-08: 0 reports for this device; peer average 5.01/device2024-09: 0 reports for this device; peer average 5.19/device2024-10: 0 reports for this device; peer average 5.7/device2024-11: 0 reports for this device; peer average 5.29/device2024-12: 0 reports for this device; peer average 4.06/device2025-01: 0 reports for this device; peer average 5.17/device2025-02: 0 reports for this device; peer average 4.46/device2025-03: 0 reports for this device; peer average 6.28/device2025-04: 0 reports for this device; peer average 4.39/device2025-05: 0 reports for this device; peer average 5.16/device2025-06: 0 reports for this device; peer average 5.08/device2025-07: 0 reports for this device; peer average 5.2/device2025-08: 0 reports for this device; peer average 5.45/device2025-09: 0 reports for this device; peer average 4.85/device2025-10: 0 reports for this device; peer average 5.14/device2025-11: 0 reports for this device; peer average 5.55/device2025-12: 0 reports for this device; peer average 5.81/device2026-01: 0 reports for this device; peer average 5.63/device2026-02: 0 reports for this device; peer average 5.98/device2026-03: 0 reports for this device; peer average 6.2/device2026-04: 0 reports for this device; peer average 5.43/device2026-05: 0 reports for this device; peer average 6.19/device2026-06: 0 reports for this device; peer average 5.56/device010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506
Monthly adverse-event reports for this device versus the product-code average.
MonthThis device (reports)Product-code average (reports per device)
2019-0106.86
2019-0204.35
2019-0304.42
2019-0406.12
2019-0504.8
2019-0605.04
2019-0705.86
2019-0806.1
2019-0904.99
2019-1005.85
2019-1105.62
2019-1205.41
2020-0104.81
2020-0205.57
2020-0305.68
2020-0403.57
2020-0503.76
2020-0605.14
2020-0706.56
2020-0804.64
2020-0905.86
2020-1005.55
2020-1105.14
2020-1205.31
2021-0105.17
2021-0204.89
2021-0305.7
2021-0405.26
2021-0505.14
2021-0605.83
2021-0704.89
2021-0804.3
2021-0904.77
2021-1006.17
2021-1105.57
2021-1205.56
2022-0105.04
2022-0205.15
2022-0305.75
2022-0404.98
2022-0505.78
2022-0606.3
2022-0704.36
2022-0814.59
2022-0905
2022-1014.76
2022-1104.61
2022-1204.04
2023-0105.52
2023-0224.59
2023-0304.87
2023-0424.78
2023-0525.14
2023-0614.78
2023-0704.82
2023-0814.44
2023-0925.23
2023-10845.1
2023-1104.38
2023-1214.79
2024-0105.54
2024-0205.46
2024-0306.16
2024-0406.04
2024-0506.28
2024-0605.39
2024-0705.05
2024-0805.01
2024-0905.19
2024-1005.7
2024-1105.29
2024-1204.06
2025-0105.17
2025-0204.46
2025-0306.28
2025-0404.39
2025-0505.16
2025-0605.08
2025-0705.2
2025-0805.45
2025-0904.85
2025-1005.14
2025-1105.55
2025-1205.81
2026-0105.63
2026-0205.98
2026-0306.2
2026-0405.43
2026-0506.19
2026-0605.56
This device (reports/month)Average device in GEI (reports/device/month)
Failure-mode summaryAI-generated

40 linked report narratives available — no AI summary generated yet.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2024-10-04Z-0366-2025Class IIOpen, ClassifiedThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuriesDistribution: Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.295,760 eachescode info di
Individual MAUDE reports
DateReport #TypeNarrativeLink basis
2023-12-210001811755-2023-00247MalfunctionAvailableexact di
2023-10-190001811755-2023-00185MalfunctionAvailableexact di
2023-10-170001811755-2023-00169MalfunctionAvailableexact di
2023-10-170001811755-2023-00146MalfunctionAvailableexact di
2023-10-170001811755-2023-00183MalfunctionAvailableexact di
2023-10-170001811755-2023-00177MalfunctionAvailableexact di
2023-10-170001811755-2023-00203MalfunctionAvailableexact di
2023-10-170001811755-2023-00144MalfunctionAvailableexact di
2023-10-170001811755-2023-00133MalfunctionAvailableexact di
2023-10-170001811755-2023-00168MalfunctionAvailableexact di
2023-10-170001811755-2023-00157MalfunctionAvailableexact di
2023-10-170001811755-2023-00217MalfunctionAvailableexact di
2023-10-170001811755-2023-00151MalfunctionAvailableexact di
2023-10-170001811755-2023-00131MalfunctionAvailableexact di
2023-10-170001811755-2023-00141MalfunctionAvailableexact di
Showing 115 of 97
Regulatory clearances

None linked.

Manufacturer compliance record
DateTypeOutcomeFacilityCitations
2026-06-18NAINo action indicatedInternational Life Sciences0
2026-02-20VAIVoluntary action indicatedStryker Puerto Rico, LLC2
2025-12-11NAINo action indicatedStryker Manufacturing Kalamazoo0
2025-05-22NAINo action indicatedStryker Macroom0
2025-03-27VAIVoluntary action indicatedStryker GmbH1
2025-03-27VAIVoluntary action indicatedStryker GmbH1
2025-02-27NAINo action indicatedStryker Leibinger GmbH & Co. KG0
2025-01-30NAINo action indicatedSerf0
2024-12-18VAIVoluntary action indicatedStryker Neurovascular1
2024-11-15VAIVoluntary action indicatedStryker Medical Division of Stryker Corporation2
2024-11-13NAINo action indicatedPOM Medical LLC0
2024-07-25VAIVoluntary action indicatedStryker Ireland, Ltd., Instruments Division1
2024-05-03NAINo action indicatedStryker Neurovascular, Inc.0
2024-04-18NAINo action indicatedPhysio-Control, Inc.0
2023-12-01VAIVoluntary action indicatedHowmedica Osteonics Corp.1
2023-05-15VAIVoluntary action indicatedLimFlow, Inc.2
2023-04-25NAINo action indicatedStryker Puerto Rico, LLC0
2023-01-24VAIVoluntary action indicatedStryker Corporation2
2023-01-24VAIVoluntary action indicatedStryker Corporation2
2022-03-08VAIVoluntary action indicatedPOM Medical LLC0
2017-04-06WLWarning LetterEntellus Medical Inc.
2015-04-08WLWarning LetterThermedx LLC
2013-11-26INJInjunctionPhysio-Control, Inc.
2013-03-06WLWarning LetterStryker Corporation
2013-03-06WLWarning LetterStryker Corporation