Adverse-event reports (MAUDE)
Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.
No adverse-event reports are linked to this device.
Recall history
| Initiated | Recall # | Class | Status | Reason | Quantity | Link basis |
|---|---|---|---|---|---|---|
| 2024-05-10 | Z-2041-2024 | Class II | Open, Classified | The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.Distribution: US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR | 153 | code info di |
| 2024-01-29 | Z-1250-2024 | Class II | Open, Classified | For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.Distribution: Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO. | 40 | code info di |
| 2021-05-21 | Z-2044-2021 | Class II | Open, Classified | The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.Distribution: US Nationwide distributions in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY. | clearance family | |
| 2018-05-30 | Z-0199-2019 | Class II | Open, Classified | When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.Distribution: Domestic: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY. International: Algeria, Australia, Bangladesh, Brazil, Canada, China, Colombia, Congo, Dominican Republic, Egypt, Guinea, Honduras, India, Indonesia, Iran, Japan, Kenya, Korea, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Nigeria, Oman, Pakistan, Paraguay, Qatar, Saudi Arabia, Singapore, South Africa, Sudan, Taiwan, Thailand, Turkey, UAE, and Vietnam | 252 | clearance family |
| 2012-10-15 | Z-1101-2013 | Class II | Terminated | It was found that there is a problem with the LCD monitor used in the console. The LCD monitor may become blank during use. Specifically, the backlight of the LCD monitor may turn off while the system is in use, causing the LCD monitor to become blank. However, the other parts of the system continue to operate normally, and the risk of a serious health hazard is therefore very low. The LCD MoniDistribution: USA Nationwide Distribution including the states of : IL, NH, MA, MI, MO, PR, NV, NM, MN, ID, NY, OH, NC, AZ, MS, LA, OR, PA, CA, and CT. | 27 units | clearance family |
Regulatory clearances
- 510(k)K122613Substantially equivalent (SESE) · 2012-12-13clearance family
- 510(k)K120638Substantially equivalent (SESE) · 2012-06-01clearance family
- 510(k)K112003Substantially equivalent (SESE) · 2011-12-02clearance family
Predicate lineage
Lineage each submission cites, walked transitively (indented = the cited device's own predicates). Extracted from public 510(k) summary text — heuristic, and includes reference devices mentioned alongside true predicates.
Manufacturer compliance record
| Date | Type | Outcome | Facility | Citations |
|---|---|---|---|---|
| 2014-09-19 | NAI | No action indicated | Canon Components, Inc. (Factory Site) | 0 |
| 2014-09-17 | NAI | No action indicated | Canon Inc. | 0 |
| 2014-09-17 | NAI | No action indicated | Canon Inc. | 0 |
Similar devices
- Achieva Philips Medical Systems Nederland B.V.
- Breast Pad Insert H SatPad™ Sat Pad Inc.
- Breast Pad Insert SatPad™ Sat Pad Inc.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
Select up to two to place side-by-side with this device.