Skip to content
Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

No adverse-event reports are linked to this device.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2025-11-06Z-0940-2026Class IIOpen, ClassifiedFollowing a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.Distribution: Domestic: WI. International: Ireland.code info di
Regulatory clearances

None linked.

Manufacturer compliance record

None in the loaded window.

Similar devices
Select up to two to place side-by-side with this device.