Adverse-event reports (MAUDE)
Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.
No adverse-event reports are linked to this device.
Recall history
| Initiated | Recall # | Class | Status | Reason | Quantity | Link basis |
|---|---|---|---|---|---|---|
| 2025-11-06 | Z-0940-2026 | Class II | Open, Classified | Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.Distribution: Domestic: WI. International: Ireland. | code info di |
Regulatory clearances
None linked.
Manufacturer compliance record
None in the loaded window.
Similar devices
- ALGO Natus Medical Denmark ApS
- 1077 AccuScreen Natus Medical Denmark ApS
- ABR Amplifier & Control service kit OTODYNAMICS LIMITED
- ABR Case service kit - Bottom OTODYNAMICS LIMITED
- ABR Case service kit - Top OTODYNAMICS LIMITED
- ABR Case service kit – latch locking OTODYNAMICS LIMITED
- ABR Case service kit – latch non locking OTODYNAMICS LIMITED
- ABR Desktop Stand/Crib Carry Crib OTODYNAMICS LIMITED
- ABR Ear Cups OTODYNAMICS LIMITED
- ABR Electrode Cable (SNAP) for OtoNova Pro OTODYNAMICS LIMITED
- ABR Electrode Cable (TAB) for OtoNova Pro OTODYNAMICS LIMITED
- ABR Electrode Cable V2 (NHSP) for Otoport V2/Pro OTODYNAMICS LIMITED
Select up to two to place side-by-side with this device.