Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.
No adverse-event reports are linked to this device.
| Initiated | Recall # | Class | Status | Reason | Quantity | Link basis |
|---|---|---|---|---|---|---|
| 2015-12-07 | Z-0467-2016 | Class II | Terminated | Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.Distribution: Nationwide including : DC, PR, Virgin Islands and Bermuda | in US 3752 units | clearance family |
Predicate lineage
Cited as predicate by: K123510, K123920, K123938, K130262, K130749, K132831, K132951, K133692, K140666, K141977, K143020, K152602, K163234, K170396, K170840, K181613, K191050, K220018
- K151579
- K153343
Cited as predicate by: K163211, K163234, K163294, K163312, K170396, K170840, K173592, K173600, K181322, K181433, K181613, K182129, K191050, K192496, K193324, K202014, K250443
- K170396
Lineage each submission cites, walked transitively (indented = the cited device's own predicates). Extracted from public 510(k) summary text — heuristic, and includes reference devices mentioned alongside true predicates.
| Date | Type | Outcome | Facility | Citations |
|---|---|---|---|---|
| 2014-09-25 | NAI | No action indicated | Siemens AG/Siemens Healthcare GmbH | 0 |
| 2011-06-09 | VAI | Voluntary action indicated | Siemens Healthcare GmbH | 1 |
| 2010-09-14 | NAI | No action indicated | Siemens AG/Siemens Healthcare GmbH | 0 |
- Achieva Philips Medical Systems Nederland B.V.
- Breast Pad Insert H SatPad™ Sat Pad Inc.
- Breast Pad Insert SatPad™ Sat Pad Inc.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.
- "MRI Integral ""T"" with Check Valve for MEDRAD® Spectris MR Injector" BAYER MEDICAL CARE INC.