Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

No adverse-event reports are linked to this device.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2015-12-07Z-0467-2016Class IITerminatedPossible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.Distribution: Nationwide including : DC, PR, Virgin Islands and Bermudain US 3752 unitsclearance family
Manufacturer compliance record
DateTypeOutcomeFacilityCitations
2014-09-25NAINo action indicatedSiemens AG/Siemens Healthcare GmbH0
2011-06-09VAIVoluntary action indicatedSiemens Healthcare GmbH1
2010-09-14NAINo action indicatedSiemens AG/Siemens Healthcare GmbH0