Unnamed device

Model RM7C · GE Healthcare Austria GmbH & Co OG · DI 00840682143271

ITXTransducer, Ultrasonic, Diagnostic · Class II · 21 CFR 892.1570

IYNSystem, Imaging, Pulsed Doppler, Ultrasonic · Class II · 21 CFR 892.1550

IYOSystem, Imaging, Pulsed Echo, Ultrasonic · Class II · 21 CFR 892.1560

UDILibrary device dossier · generated 2026-09-16 · sources: FDA GUDID, openFDA (recalls, MAUDE, 510(k)/PMA), FDA Data Dashboard (inspections, compliance actions), 510(k) summary documents.

Recall history (1)

InitiatedRecall #ClassStatusReasonLink basis
2019-07-02Z-1964-2019Class IITerminatedPotential for a transducer mis-alignment in certain transvaginal probes.k number procode

Adverse-event reports — MAUDE (0)

Caveat (required reading): MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line is a reporting baseline (average device in the same product code), not a risk measure.

5914192019-01: 0 reports for this device; peer average 3.15/device2019-02: 0 reports for this device; peer average 5.46/device2019-03: 0 reports for this device; peer average 5.81/device2019-04: 0 reports for this device; peer average 3.3/device2019-05: 0 reports for this device; peer average 5.27/device2019-06: 0 reports for this device; peer average 6.28/device2019-07: 0 reports for this device; peer average 5.59/device2019-08: 0 reports for this device; peer average 7.45/device2019-09: 0 reports for this device; peer average 3.92/device2019-10: 0 reports for this device; peer average 3.57/device2019-11: 0 reports for this device; peer average 4.04/device2019-12: 0 reports for this device; peer average 4.84/device2020-01: 0 reports for this device; peer average 2.75/device2020-02: 0 reports for this device; peer average 5.41/device2020-03: 0 reports for this device; peer average 5.5/device2020-04: 0 reports for this device; peer average 5.17/device2020-05: 0 reports for this device; peer average 2.89/device2020-06: 0 reports for this device; peer average 3.16/device2020-07: 0 reports for this device; peer average 4.32/device2020-08: 0 reports for this device; peer average 2.93/device2020-09: 0 reports for this device; peer average 2.07/device2020-10: 0 reports for this device; peer average 3.79/device2020-11: 0 reports for this device; peer average 4.01/device2020-12: 0 reports for this device; peer average 5.87/device2021-01: 0 reports for this device; peer average 6.53/device2021-02: 0 reports for this device; peer average 4.61/device2021-03: 0 reports for this device; peer average 7.44/device2021-04: 0 reports for this device; peer average 4.17/device2021-05: 0 reports for this device; peer average 5.98/device2021-06: 0 reports for this device; peer average 6.44/device2021-07: 0 reports for this device; peer average 8.34/device2021-08: 0 reports for this device; peer average 7.22/device2021-09: 0 reports for this device; peer average 9.35/device2021-10: 0 reports for this device; peer average 5.7/device2021-11: 0 reports for this device; peer average 6.88/device2021-12: 0 reports for this device; peer average 9.64/device2022-01: 0 reports for this device; peer average 6.49/device2022-02: 0 reports for this device; peer average 8.37/device2022-03: 0 reports for this device; peer average 7.24/device2022-04: 0 reports for this device; peer average 8.11/device2022-05: 0 reports for this device; peer average 9.67/device2022-06: 0 reports for this device; peer average 7.56/device2022-07: 0 reports for this device; peer average 7.46/device2022-08: 0 reports for this device; peer average 5.42/device2022-09: 0 reports for this device; peer average 4.73/device2022-10: 0 reports for this device; peer average 4.62/device2022-11: 0 reports for this device; peer average 4.41/device2022-12: 0 reports for this device; peer average 5.16/device2023-01: 0 reports for this device; peer average 5.01/device2023-02: 0 reports for this device; peer average 4.61/device2023-03: 0 reports for this device; peer average 5.75/device2023-04: 0 reports for this device; peer average 3.78/device2023-05: 0 reports for this device; peer average 6.15/device2023-06: 0 reports for this device; peer average 4.06/device2023-07: 0 reports for this device; peer average 4.25/device2023-08: 0 reports for this device; peer average 5.94/device2023-09: 0 reports for this device; peer average 5.7/device2023-10: 0 reports for this device; peer average 6.69/device2023-11: 0 reports for this device; peer average 5.36/device2023-12: 0 reports for this device; peer average 5.47/device2024-01: 0 reports for this device; peer average 6.28/device2024-02: 0 reports for this device; peer average 4.31/device2024-03: 0 reports for this device; peer average 5.97/device2024-04: 0 reports for this device; peer average 5.86/device2024-05: 0 reports for this device; peer average 4.16/device2024-06: 0 reports for this device; peer average 5.85/device2024-07: 0 reports for this device; peer average 4.97/device2024-08: 0 reports for this device; peer average 5.4/device2024-09: 0 reports for this device; peer average 5.79/device2024-10: 0 reports for this device; peer average 6.4/device2024-11: 0 reports for this device; peer average 4.03/device2024-12: 0 reports for this device; peer average 11.27/device2025-01: 0 reports for this device; peer average 4.34/device2025-02: 0 reports for this device; peer average 4.03/device2025-03: 0 reports for this device; peer average 5.83/device2025-04: 0 reports for this device; peer average 3.25/device2025-05: 0 reports for this device; peer average 3.93/device2025-06: 0 reports for this device; peer average 4.11/device2025-07: 0 reports for this device; peer average 4.1/device2025-08: 0 reports for this device; peer average 5.45/device2025-09: 0 reports for this device; peer average 6.52/device2025-10: 0 reports for this device; peer average 9.71/device2025-11: 0 reports for this device; peer average 12.53/device2025-12: 0 reports for this device; peer average 18.81/device2026-01: 0 reports for this device; peer average 11.99/device2026-02: 0 reports for this device; peer average 6.63/device2026-03: 0 reports for this device; peer average 8.63/device2026-04: 0 reports for this device; peer average 11.88/device2026-05: 0 reports for this device; peer average 16.78/device2026-06: 0 reports for this device; peer average 15.26/device010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506
Monthly adverse-event reports for this device versus the product-code average.
MonthThis device (reports)Product-code average (reports per device)
2019-0103.15
2019-0205.46
2019-0305.81
2019-0403.3
2019-0505.27
2019-0606.28
2019-0705.59
2019-0807.45
2019-0903.92
2019-1003.57
2019-1104.04
2019-1204.84
2020-0102.75
2020-0205.41
2020-0305.5
2020-0405.17
2020-0502.89
2020-0603.16
2020-0704.32
2020-0802.93
2020-0902.07
2020-1003.79
2020-1104.01
2020-1205.87
2021-0106.53
2021-0204.61
2021-0307.44
2021-0404.17
2021-0505.98
2021-0606.44
2021-0708.34
2021-0807.22
2021-0909.35
2021-1005.7
2021-1106.88
2021-1209.64
2022-0106.49
2022-0208.37
2022-0307.24
2022-0408.11
2022-0509.67
2022-0607.56
2022-0707.46
2022-0805.42
2022-0904.73
2022-1004.62
2022-1104.41
2022-1205.16
2023-0105.01
2023-0204.61
2023-0305.75
2023-0403.78
2023-0506.15
2023-0604.06
2023-0704.25
2023-0805.94
2023-0905.7
2023-1006.69
2023-1105.36
2023-1205.47
2024-0106.28
2024-0204.31
2024-0305.97
2024-0405.86
2024-0504.16
2024-0605.85
2024-0704.97
2024-0805.4
2024-0905.79
2024-1006.4
2024-1104.03
2024-12011.27
2025-0104.34
2025-0204.03
2025-0305.83
2025-0403.25
2025-0503.93
2025-0604.11
2025-0704.1
2025-0805.45
2025-0906.52
2025-1009.71
2025-11012.53
2025-12018.81
2026-01011.99
2026-0206.63
2026-0308.63
2026-04011.88
2026-05016.78
2026-06015.26
This device (reports/month)Average device in ITX, IYN, IYO (reports/device/month)

Regulatory clearances

  • K181985 SESE 2018-11-15 (k number procode)

Predicate lineage

K181985 — cites (transitively): K172342 (d1), K173555 (d1), K163077 (d2), K170445 (d2), K152309 (d3), K150087 (d3), K152195 (d3), K160182 (d3), K160184 (d3), K160524 (d3), K161047 (d3), K131514 (d4), K142160 (d4), K142323 (d4), K131527 (d4), K141261 (d4), K143452 (d4), K141639 (d4), K142472 (d4), K152567 (d4), K150949 (d4), K133034 (d4), K133533 (d4), K151028 (d4), K160277 (d4), K031663 (d5), K120201 (d5), K120221 (d5), K123564 (d5), K101874 (d5), K120741 (d5), K132913 (d5), K132274 (d5), K121063 (d5), K141093 (d5), K123806 (d5), K142517 (d5), K113690 (d5), K141768 (d5), K152758 (d5), K073297 (d6), K092271 (d6), K102388 (d6), K053435 (d6), K103649 (d6), K071985 (d6), K082374 (d6), K092079 (d6), K111582 (d6), K122327 (d6), K122921 (d6), K112255 (d6), K121329 (d6), K060907 (d6), K131267 (d6), K123209 (d6), K133761 (d6), K133431 (d6), K142443 (d6), K033139 (d7), K091970 (d7), K102041 (d7), K072786 (d7), K093563 (d7), K080555 (d7), K123992 (d7), K071204 (d7), K092619 (d7), K132942 (d7), K021966 (d8), K023877 (d8), K030455 (d8), K042540 (d8), K103629 (d8), K121422 (d8), K100402 (d8) — cited as predicate by K192152, K192159, K192505, K210699, K240115, K240684

Lineage extracted from public 510(k) summary text (heuristic).

Manufacturer compliance record — GE Healthcare Austria GmbH & Co OG

  • NAI 2013-11-20 GE HEALTHCARE AUSTRIA GMBH & CO OG · 0 citations