Discovery

Model NMCT 670 CZT · GE Medical Systems Israel, Functional Imaging · DI 00840682139380

JAKSystem, X-Ray, Tomography, Computed · Class II · 21 CFR 892.1750

KPSSystem, Tomography, Computed, Emission · Class II · 21 CFR 892.1200

UDILibrary device dossier · generated 2026-09-15 · sources: FDA GUDID, openFDA (recalls, MAUDE, 510(k)/PMA), FDA Data Dashboard (inspections, compliance actions), 510(k) summary documents.

Recall history (6)

InitiatedRecall #ClassStatusReasonLink basis
2022-12-19Z-1057-2023Class IOpen, ClassifiedA mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.k number procode
2020-12-10Z-0855-2021Class IIOpen, ClassifiedGE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.k number procode
2020-08-14Z-2993-2020Class IITerminatedRotor bearing screws were found loose on detector.k number procode
2019-03-20Z-1258-2019Class IITerminatedDuring installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.k number procode
2019-03-20Z-1264-2019Class IITerminatedDuring installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.k number procode
2017-05-19Z-2195-2017Class IITerminatedIt was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.k number procode

Adverse-event reports — MAUDE (0)

Caveat (required reading): MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line is a reporting baseline (average device in the same product code), not a risk measure.

23572019-01: 0 reports for this device; peer average 5.1/device2019-02: 0 reports for this device; peer average 3.3/device2019-03: 0 reports for this device; peer average 3.15/device2019-04: 0 reports for this device; peer average 5.74/device2019-05: 0 reports for this device; peer average 3.31/device2019-06: 0 reports for this device; peer average 2.8/device2019-07: 0 reports for this device; peer average 2.58/device2019-08: 0 reports for this device; peer average 2.38/device2019-09: 0 reports for this device; peer average 2.14/device2019-10: 0 reports for this device; peer average 2.25/device2019-11: 0 reports for this device; peer average 1.86/device2019-12: 0 reports for this device; peer average 5.33/device2020-01: 0 reports for this device; peer average 1.92/device2020-02: 0 reports for this device; peer average 2.75/device2020-03: 0 reports for this device; peer average 2.17/device2020-04: 0 reports for this device; peer average 3.67/device2020-05: 0 reports for this device; peer average 2/device2020-06: 0 reports for this device; peer average 2.44/device2020-07: 0 reports for this device; peer average 2.25/device2020-08: 0 reports for this device; peer average 1.74/device2020-09: 0 reports for this device; peer average 1.95/device2020-10: 0 reports for this device; peer average 2.05/device2020-11: 0 reports for this device; peer average 2.24/device2020-12: 0 reports for this device; peer average 2.4/device2021-01: 0 reports for this device; peer average 1.47/device2021-02: 0 reports for this device; peer average 1.82/device2021-03: 0 reports for this device; peer average 2.43/device2021-04: 0 reports for this device; peer average 2.74/device2021-05: 0 reports for this device; peer average 3.77/device2021-06: 0 reports for this device; peer average 4.55/device2021-07: 0 reports for this device; peer average 3.07/device2021-08: 0 reports for this device; peer average 2.53/device2021-09: 0 reports for this device; peer average 2.3/device2021-10: 0 reports for this device; peer average 1.67/device2021-11: 0 reports for this device; peer average 2.62/device2021-12: 0 reports for this device; peer average 2.75/device2022-01: 0 reports for this device; peer average 1.89/device2022-02: 0 reports for this device; peer average 2.47/device2022-03: 0 reports for this device; peer average 2.43/device2022-04: 0 reports for this device; peer average 2.45/device2022-05: 0 reports for this device; peer average 1.73/device2022-06: 0 reports for this device; peer average 1.88/device2022-07: 0 reports for this device; peer average 3.53/device2022-08: 0 reports for this device; peer average 1.86/device2022-09: 0 reports for this device; peer average 2.81/device2022-10: 0 reports for this device; peer average 2.09/device2022-11: 0 reports for this device; peer average 2.07/device2022-12: 0 reports for this device; peer average 1.65/device2023-01: 0 reports for this device; peer average 2.31/device2023-02: 0 reports for this device; peer average 1.9/device2023-03: 0 reports for this device; peer average 2.63/device2023-04: 0 reports for this device; peer average 2.17/device2023-05: 0 reports for this device; peer average 2.57/device2023-06: 0 reports for this device; peer average 6.62/device2023-07: 0 reports for this device; peer average 2.1/device2023-08: 0 reports for this device; peer average 3.42/device2023-09: 0 reports for this device; peer average 2.95/device2023-10: 0 reports for this device; peer average 5.89/device2023-11: 0 reports for this device; peer average 2.58/device2023-12: 0 reports for this device; peer average 2.74/device2024-01: 0 reports for this device; peer average 1.9/device2024-02: 0 reports for this device; peer average 1.88/device2024-03: 0 reports for this device; peer average 1.85/device2024-04: 0 reports for this device; peer average 1.72/device2024-05: 0 reports for this device; peer average 2.24/device2024-06: 0 reports for this device; peer average 3.08/device2024-07: 0 reports for this device; peer average 1.88/device2024-08: 0 reports for this device; peer average 1.33/device2024-09: 0 reports for this device; peer average 1.17/device2024-10: 0 reports for this device; peer average 1.95/device2024-11: 0 reports for this device; peer average 1.45/device2024-12: 0 reports for this device; peer average 1.69/device2025-01: 0 reports for this device; peer average 1.59/device2025-02: 0 reports for this device; peer average 1.58/device2025-03: 0 reports for this device; peer average 1.39/device2025-04: 0 reports for this device; peer average 1.86/device2025-05: 0 reports for this device; peer average 2.5/device2025-06: 0 reports for this device; peer average 1.5/device2025-07: 0 reports for this device; peer average 1.5/device2025-08: 0 reports for this device; peer average 1.5/device2025-09: 0 reports for this device; peer average 1.67/device2025-10: 0 reports for this device; peer average 1.08/device2025-11: 0 reports for this device; peer average 1.28/device2025-12: 0 reports for this device; peer average 2.33/device2026-01: 0 reports for this device; peer average 2.22/device2026-02: 0 reports for this device; peer average 1.68/device2026-03: 0 reports for this device; peer average 1.45/device2026-04: 0 reports for this device; peer average 2/device2026-05: 0 reports for this device; peer average 2.19/device2026-06: 0 reports for this device; peer average 1.55/device010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506070809101112010203040506
Monthly adverse-event reports for this device versus the product-code average.
MonthThis device (reports)Product-code average (reports per device)
2019-0105.1
2019-0203.3
2019-0303.15
2019-0405.74
2019-0503.31
2019-0602.8
2019-0702.58
2019-0802.38
2019-0902.14
2019-1002.25
2019-1101.86
2019-1205.33
2020-0101.92
2020-0202.75
2020-0302.17
2020-0403.67
2020-0502
2020-0602.44
2020-0702.25
2020-0801.74
2020-0901.95
2020-1002.05
2020-1102.24
2020-1202.4
2021-0101.47
2021-0201.82
2021-0302.43
2021-0402.74
2021-0503.77
2021-0604.55
2021-0703.07
2021-0802.53
2021-0902.3
2021-1001.67
2021-1102.62
2021-1202.75
2022-0101.89
2022-0202.47
2022-0302.43
2022-0402.45
2022-0501.73
2022-0601.88
2022-0703.53
2022-0801.86
2022-0902.81
2022-1002.09
2022-1102.07
2022-1201.65
2023-0102.31
2023-0201.9
2023-0302.63
2023-0402.17
2023-0502.57
2023-0606.62
2023-0702.1
2023-0803.42
2023-0902.95
2023-1005.89
2023-1102.58
2023-1202.74
2024-0101.9
2024-0201.88
2024-0301.85
2024-0401.72
2024-0502.24
2024-0603.08
2024-0701.88
2024-0801.33
2024-0901.17
2024-1001.95
2024-1101.45
2024-1201.69
2025-0101.59
2025-0201.58
2025-0301.39
2025-0401.86
2025-0502.5
2025-0601.5
2025-0701.5
2025-0801.5
2025-0901.67
2025-1001.08
2025-1101.28
2025-1202.33
2026-0102.22
2026-0201.68
2026-0301.45
2026-0402
2026-0502.19
2026-0601.55
This device (reports/month)Average device in JAK, KPS (reports/device/month)

Regulatory clearances

  • K153402 SESE 2016-01-21 (k number procode)

Predicate lineage

K153402 — cites (transitively): K093514 (d1), K102231 (d1), K022960 (d2), K993813 (d2) — cited as predicate by K161740, K182484, K210173

Lineage extracted from public 510(k) summary text (heuristic).

Manufacturer compliance record — GE Medical Systems Israel, Functional Imaging

  • NAI 2010-01-12 GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · 0 citations