Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

No adverse-event reports are linked to this device.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2018-09-04Z-0095-2019Class IIOpen, ClassifiedWhen multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.Distribution: Worldwide Distribution-US Nationwide59,000 unitsclearance family
2018-09-04Z-0096-2019Class IIOpen, ClassifiedWhen multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.Distribution: Worldwide Distribution-US Nationwide47,937 unitsclearance family
2015-05-04Z-1606-2015Class IITerminatedThe heart rate could be calculated from pacer pulses without indication that pacemaker detection is OFF in some combinations of Bx50 monitors, a PDM, and CIC/CSCS. Then the monitor does not show the Pacer Off message and there may not be an alarm for asystole. Undetected asystole could result in irreversible changes in the patients condition and delayed or missed life sustaining patient treatmentDistribution: Worldwide Distribution - US including CO, MI, WI and Internationally to NETHERLANDS, FINLAND, and BELGIUM.110 (83 units US, 27 units OUS)clearance family
Regulatory clearances
  • 510(k)K132533Substantially equivalent (SESE) · 2013-12-04clearance family

Predicate lineage

Lineage each submission cites, walked transitively (indented = the cited device's own predicates). Extracted from public 510(k) summary text — heuristic, and includes reference devices mentioned alongside true predicates.

Manufacturer compliance record
DateTypeOutcomeFacilityCitations
2012-02-02VAIVoluntary action indicatedGE Healthcare Finland Oy1