Adverse-event reports (MAUDE)

Read this first: MAUDE report counts cannot be used to compute incidence rates or compare device safety. Reporting is incomplete, duplicated, and spikes with publicity; there is no denominator of devices in use. The peer line shows reporting volume for the average device in the same FDA product code — a reporting baseline, not a risk measure.

1 linked reportInjury 1
01112026-04: 1 reports for this device; peer average 1/device04
Monthly adverse-event reports for this device versus the product-code average.
MonthThis device (reports)Product-code average (reports per device)
2026-0411
This device (reports/month)Average device in OUO (reports/device/month)
Failure-mode summaryAI-generated

1 linked report narrative available — no AI summary generated yet.

Recall history
InitiatedRecall #ClassStatusReasonQuantityLink basis
2025-06-20Z-2309-2025Class IIOpen, ClassifiedGE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.Distribution: Worldwide distribution.97 unitscode info di
2025-05-23Z-2133-2025Class IIOpen, ClassifiedGE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.Distribution: Worldwide distribution - US Nationwide and the country of China.N/Acode info di
2024-06-14Z-2466-2024Class IIOpen, ClassifiedGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.Distribution: Worldwide distribution.123 unitsclearance family
2021-12-23Z-0595-2022Class IIOpen, ClassifiedThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.Distribution: Distribution was nationwide, including Puerto Rico and the Virgin Islands. There was also government and military distribution. Foreign distribution was made to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaidjan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Republic of Congo, Croatia, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Equatorial Guinea, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrhyzstan, Latvia, Lebanon, Libyan Arabic Jamahiriya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Mali, Malta, Mauritania, Mexico, Republic of Moldova, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Nordic-HQ, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Zambia, and Zimbabwe.17,228 devices total in this field correctionclearance family
2021-06-04Z-2057-2021Class IIOpen, ClassifiedGE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.Distribution: Worldwide distributions.71 devicesclearance family
2020-09-08Z-0131-2021Class IITerminatedPotential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.Distribution: Worldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.N/Aclearance family
2015-08-31Z-0260-2016Class IITerminatedWhen performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.Distribution: Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.17clearance family
Individual MAUDE reports
DateReport #TypeNarrativeLink basis
2026-04-152183553-2026-00002InjuryAvailableexact di
Regulatory clearances
  • 510(k)K163619Substantially equivalent (SESE) · 2017-02-14clearance family
  • 510(k)K142098Substantially equivalent (SESE) · 2014-11-24clearance family

Predicate lineage

Lineage each submission cites, walked transitively (indented = the cited device's own predicates). Extracted from public 510(k) summary text — heuristic, and includes reference devices mentioned alongside true predicates.

Manufacturer compliance record

None in the loaded window.